The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter

<h3>Purpose</h3><p dir="ltr">To investigate the operation principles of the automatic tube current modulation (ATCM) of a modern GE healthcare CT scanner, and the impact of related settings on image quality and patient dose.</p><h3>Material & Methods</h...

وصف كامل

محفوظ في:
التفاصيل البيبلوغرافية
المؤلف الرئيسي: Ioannis A. Tsalafoutas (14776939) (author)
مؤلفون آخرون: Shady AlKhazzam (14776942) (author), Mohammed Hassan Kharita (14776945) (author)
منشور في: 2024
الموضوعات:
الوسوم: إضافة وسم
لا توجد وسوم, كن أول من يضع وسما على هذه التسجيلة!
الوصف
الملخص:<h3>Purpose</h3><p dir="ltr">To investigate the operation principles of the automatic tube current modulation (ATCM) of a modern GE healthcare CT scanner, and the impact of related settings on image quality and patient dose.</p><h3>Material & Methods</h3><p dir="ltr">A dedicated phantom (Mercury 4.0) was scanned using two of the most frequently used clinical scanning protocols (chest and abdomen‐pelvis). The preset protocol settings were used as starting points (reference conditions). Scan direction, scan mode (helical vs. axial), total beam width, tube potential (kVp), and ATCM settings were then modified individually to understand their impact on radiation dose and image quality. Regarding the ATCM settings, the SmartmA minimum and maximum mA limits, and the noise index (NI) values were varied. As surrogates of patient dose, the CTDIvol and DLP values of each scan were used. As surrogates of image quality were used the image noise and the detectability index (d’) of five different materials (air, solid water, polystyrene, iodine, and bone) embedded in the Mercury phantom calculated with the ImQuest software.</p><h3>Results</h3><p dir="ltr">The scanning direction did not have any effect on ATCM curves, unlike what has been observed in CT scanners from other manufacturers. Total beam width does matter, however, the SmartmA limit settings and kVp selection had the greatest impact on image quality and dose. It was seen that improper minimum mA limit settings practically invalidated the ATCM operation. In contrast, when full modulation was allowed without restrictions, noise standard deviation, and detectability index became much more consistent across the wide range of phantom diameters. For lower kVp settings an impressive dose reduction was observed that requires further investigation.</p><h3>Conclusion</h3><p dir="ltr">SmartmA is a tool that if not properly used may increase the patient doses considerably. Therefore, its settings should be carefully adjusted for each preset different clinical protocol.</p><h2>Other Information</h2><p dir="ltr">Published in: Journal of Applied Clinical Medical Physics<br>License: <a href="http://creativecommons.org/licenses/by/4.0/" target="_blank">http://creativecommons.org/licenses/by/4.0/</a><br>See article on publisher's website: <a href="https://dx.doi.org/10.1002/acm2.14356" target="_blank">https://dx.doi.org/10.1002/acm2.14356</a></p>