Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application

<h3>Purpose</h3><p dir="ltr">This study discusses the measurement of dose in clinical commissioning tests described in IAEA-TECDOC-1583. It explores the application of Monte Carlo (MC) modelled medium dependency correction factors (K<sub>med</sub>) for accurat...

Full description

Saved in:
Bibliographic Details
Main Author: Aram Rostami (22045274) (author)
Other Authors: Abdul Sattar Khalid (22046720) (author), Hamed Ghaffari (5857139) (author), Satheesh Prasad Paloor (22045277) (author), Bevan Orville Peltier (22046723) (author), Rabih Hammoud (14776912) (author), Shihab Abdelrahman (22046726) (author)
Published: 2024
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1864513541494538240
author Aram Rostami (22045274)
author2 Abdul Sattar Khalid (22046720)
Hamed Ghaffari (5857139)
Satheesh Prasad Paloor (22045277)
Bevan Orville Peltier (22046723)
Rabih Hammoud (14776912)
Shihab Abdelrahman (22046726)
author2_role author
author
author
author
author
author
author_facet Aram Rostami (22045274)
Abdul Sattar Khalid (22046720)
Hamed Ghaffari (5857139)
Satheesh Prasad Paloor (22045277)
Bevan Orville Peltier (22046723)
Rabih Hammoud (14776912)
Shihab Abdelrahman (22046726)
author_role author
dc.creator.none.fl_str_mv Aram Rostami (22045274)
Abdul Sattar Khalid (22046720)
Hamed Ghaffari (5857139)
Satheesh Prasad Paloor (22045277)
Bevan Orville Peltier (22046723)
Rabih Hammoud (14776912)
Shihab Abdelrahman (22046726)
dc.date.none.fl_str_mv 2024-06-03T03:00:00Z
dc.identifier.none.fl_str_mv 10.1016/j.ejmp.2024.103390
dc.relation.none.fl_str_mv https://figshare.com/articles/journal_contribution/Assessment_of_four_dose_calculation_algorithms_using_IAEA-TECDOC-1583_with_medium_dependency_correction_factor_K_sub_med_sub_application/29899823
dc.rights.none.fl_str_mv CC BY 4.0
info:eu-repo/semantics/openAccess
dc.subject.none.fl_str_mv Biomedical and clinical sciences
Clinical sciences
Medical physiology
Dose calculation algorithms
IAEA-TECDOC-1583
Medium dependency correction factor
Monte Carlo
CIRS phantom
dc.title.none.fl_str_mv Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
dc.type.none.fl_str_mv Text
Journal contribution
info:eu-repo/semantics/publishedVersion
text
contribution to journal
description <h3>Purpose</h3><p dir="ltr">This study discusses the measurement of dose in clinical commissioning tests described in IAEA-TECDOC-1583. It explores the application of Monte Carlo (MC) modelled medium dependency correction factors (K<sub>med</sub>) for accurate dose measurement in bone and lung materials using the CIRS phantom. </p><h3>Methods</h3><p dir="ltr">BEAMnrc codes simulate radiation sources and model radiation transport for 6 MV and 15 MV photon beams. CT images of the CIRS phantom are converted to an MC compatible phantom. The PTW 30013 farmer chamber measures doses within modeled CIRS phantom. K<sub>med</sub> are determined by averaging values from four central voxels within the sensitive volume of the farmer chamber. K<sub>med</sub> is calculated for D<sub>m.m</sub> and D<sub>w.w </sub>algorithm types in bone and lung media for both photon beams. </p><h3>Results</h3><p dir="ltr">Average modelled correction factors for D<sub>m.m</sub>calculations using the farmer chamber are 0.976 (±0.1 %) for 6 MV and 0.979 (±0.1 %) for 15 MV in bone media. Correspondingly, correction factors for Dw.w calculations are 0.99 (±0.3 %) and 0.992 (±0.4 %), respectively. For lung media, average correction factors for D<sub>m.m</sub> calculations are 1.02 (±0.3 %) for 6 MV and 1.022 (±0.4 %) for 15 MV. Correspondingly, correction factors for D<sub>w.w </sub>calculations are 1.01 (±0.3 %) and 1.012 (±0.2 %), respectively. </p><h3>Conclusions</h3><p dir="ltr">This study highlights the significant impact of applying K<sub>med</sub> on dose differences between measurement and calculation during the dose audit process.</p><h2>Other Information</h2><p dir="ltr">Published in: Physica Medica<br>License: <a href="http://creativecommons.org/licenses/by/4.0/" target="_blank">http://creativecommons.org/licenses/by/4.0/</a><br>See article on publisher's website: <a href="https://dx.doi.org/10.1016/j.ejmp.2024.103390" target="_blank">https://dx.doi.org/10.1016/j.ejmp.2024.103390</a></p>
eu_rights_str_mv openAccess
id Manara2_e05a4a8e6fd1ad988178a3545eb17e92
identifier_str_mv 10.1016/j.ejmp.2024.103390
network_acronym_str Manara2
network_name_str Manara2
oai_identifier_str oai:figshare.com:article/29899823
publishDate 2024
repository.mail.fl_str_mv
repository.name.fl_str_mv
repository_id_str
rights_invalid_str_mv CC BY 4.0
spelling Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) applicationAram Rostami (22045274)Abdul Sattar Khalid (22046720)Hamed Ghaffari (5857139)Satheesh Prasad Paloor (22045277)Bevan Orville Peltier (22046723)Rabih Hammoud (14776912)Shihab Abdelrahman (22046726)Biomedical and clinical sciencesClinical sciencesMedical physiologyDose calculation algorithmsIAEA-TECDOC-1583Medium dependency correction factorMonte CarloCIRS phantom<h3>Purpose</h3><p dir="ltr">This study discusses the measurement of dose in clinical commissioning tests described in IAEA-TECDOC-1583. It explores the application of Monte Carlo (MC) modelled medium dependency correction factors (K<sub>med</sub>) for accurate dose measurement in bone and lung materials using the CIRS phantom. </p><h3>Methods</h3><p dir="ltr">BEAMnrc codes simulate radiation sources and model radiation transport for 6 MV and 15 MV photon beams. CT images of the CIRS phantom are converted to an MC compatible phantom. The PTW 30013 farmer chamber measures doses within modeled CIRS phantom. K<sub>med</sub> are determined by averaging values from four central voxels within the sensitive volume of the farmer chamber. K<sub>med</sub> is calculated for D<sub>m.m</sub> and D<sub>w.w </sub>algorithm types in bone and lung media for both photon beams. </p><h3>Results</h3><p dir="ltr">Average modelled correction factors for D<sub>m.m</sub>calculations using the farmer chamber are 0.976 (±0.1 %) for 6 MV and 0.979 (±0.1 %) for 15 MV in bone media. Correspondingly, correction factors for Dw.w calculations are 0.99 (±0.3 %) and 0.992 (±0.4 %), respectively. For lung media, average correction factors for D<sub>m.m</sub> calculations are 1.02 (±0.3 %) for 6 MV and 1.022 (±0.4 %) for 15 MV. Correspondingly, correction factors for D<sub>w.w </sub>calculations are 1.01 (±0.3 %) and 1.012 (±0.2 %), respectively. </p><h3>Conclusions</h3><p dir="ltr">This study highlights the significant impact of applying K<sub>med</sub> on dose differences between measurement and calculation during the dose audit process.</p><h2>Other Information</h2><p dir="ltr">Published in: Physica Medica<br>License: <a href="http://creativecommons.org/licenses/by/4.0/" target="_blank">http://creativecommons.org/licenses/by/4.0/</a><br>See article on publisher's website: <a href="https://dx.doi.org/10.1016/j.ejmp.2024.103390" target="_blank">https://dx.doi.org/10.1016/j.ejmp.2024.103390</a></p>2024-06-03T03:00:00ZTextJournal contributioninfo:eu-repo/semantics/publishedVersiontextcontribution to journal10.1016/j.ejmp.2024.103390https://figshare.com/articles/journal_contribution/Assessment_of_four_dose_calculation_algorithms_using_IAEA-TECDOC-1583_with_medium_dependency_correction_factor_K_sub_med_sub_application/29899823CC BY 4.0info:eu-repo/semantics/openAccessoai:figshare.com:article/298998232024-06-03T03:00:00Z
spellingShingle Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
Aram Rostami (22045274)
Biomedical and clinical sciences
Clinical sciences
Medical physiology
Dose calculation algorithms
IAEA-TECDOC-1583
Medium dependency correction factor
Monte Carlo
CIRS phantom
status_str publishedVersion
title Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
title_full Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
title_fullStr Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
title_full_unstemmed Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
title_short Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
title_sort Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K<sub>med</sub>) application
topic Biomedical and clinical sciences
Clinical sciences
Medical physiology
Dose calculation algorithms
IAEA-TECDOC-1583
Medium dependency correction factor
Monte Carlo
CIRS phantom